The model 17405 provides a rapid and accurate determination of lung Residual Volume for the purpose of body composition analysis. It is based on our new "Tri-Gas" analyzer and uses the method of nitrogen washout.
This technique, when combined with hydrostatic weighing, is considered to be the Gold Standard in body composition determination. Perfect for accurate Residual Volume and FRC Measurements.
May also be used for Total Lung Volume using larger rebreathing balloon.
Tri-Gas Analyzer
- Gas Sampling Rate: 300 ml/min.
- Operating Range: 0 - 100% Nitrogen.
- Display Resolution: 0.1%.
- Display Linearity: 0.5% of reading from 0 - 100% N2.
- Rise Time: 10% to 90% approximately 5 sec.
- Fall Time: 90% to 10% approximately 5 sec.
- Rear Panel Analog Outputs: Separate BNC connectors for O2, N2, CO2.
Residual Volume Test
- Duration of test: 3 - 4 minutes.
- Overall Accuracy: +/-5% Menu-driven user interface Backlit LCD display.
- Computer Interface: USB.
- Auxiliary Connector: analog out.
Weight & Dimensions
- Electronics Unit Weight: Approx 4 pounds.
- Electronics Unit Dimensions: 8.25" W X 1.75" H X 10" D.
- Overall cart Size: 19" W x 44" H x 23" D.
Included Accessories: Computer interface, 2 multi-function valves, mouthpiece, filter, power cable, sample line tubing and instruction manual.