SunTech Tango+ Stress BP
Product Features
SunTech Tango+ Stress BP is a non-invasive multi-parameter monitor measuring blood pressure and oxygen saturation for use in cardiac or exercise stress testing. It measures and displays an adult patient’s systolic and diastolic blood pressure and the percent oxygen saturation of arterial blood.
Product Specification
Blood Pressure Measurement | ||
Measurement: |
Auscultatory, using R-wave gating and k-sound analysis, during all static and active phases of stress testing. Systolic pressures correlate to a K-1 Korotkoff sound. Diastolic pressures correlate to K-5 Korotkoff sound. |
|
Range: |
Pressure: |
Heart Rate: 40-200 BPM (beats per minute) |
Accuracy: |
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff /stethoscope auscultation method, within the limits prescribed by the American National Standard, Manual, electronic or automated sphygmomanometers. |
|
Conditions for Use: |
Operating: |
|
Power: |
External power supply, use only SunTech part number 19-0012-00. Input: 100-240 VAC @ 1.6A max, 50-60 Hz. Output +9VDC @ 5A IEC 320 type input connector. |
|
Calibration: |
The accuracy of cuff -pressure transducers/indicators should be verified annually. |
|
Safety Systems: |
ndependent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 300 mmHg (+20/-10mmHg). Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds. |
|
Dimensions: |
Size: 22.8 cm x 15.9 cm x 9.7 cm (9.0” x 6.3’’ x 4.1’’) |
|
Standards: |
UL60601-1, CAN/CSA C22.2 601-1 |
|
Classifications: |
Equipment Classification: Class II; Mode of Operation: Continuous |
|
Pulse Oximetry |
||
Accuracy: |
70 - 100% ± 2 digits (± 1 Standard Deviation*). |
|
* Standard Deviation is a statistical measure up to 32% of the readings may fall outside those limits. |